EPA1664—己烷提取物的检测

点击次数:4853 发布时间:2013/6/23
提 供 商: 广州分分快三下载科学仪器有限公司 资料大小:
图片类型: 下载次数: 642
资料类型: JPG 浏览次数: 4853
相关产品:
详细介绍: 文件下载    

 1) Assemble

a) Assemble cartridge adapters on the vacuum manifold

b) Place ECUNIOGXF cartridge(s) on each vacuum station needed

c) Connect the manifold to a suitable trap and attach the trap to a vacuum system capable of

attaining a minimum of 25” Hg (635 mm) of vacuum

d) Insert waste collection vial in manifold (optional)

2) Prepare Water Sample

a) Adjust the pH of the sample to less than 2 by adding 5 mL of 6N HCl or 2.5 mL of concentrated

H2SO4

b) If acid was added to the sample in the field do not add more unless the pH is greater than 2

Gloves are recommended as skin oils may affect final sample weight

3) Condition the Cartridge

a) Rinse sides of cartridge and bottle holder with 10 mL of hexane

b) Allow cartridge to soak for 1 minute

c) Draw the hexane through the cartridge using vacuum

d) Discard the hexane from waste vial

e) Draw full vacuum through the cartridge for 2 minutes to dry

f) Add 10 mL of methanol to the cartridge

g) Slowly draw the methanol through leaving a layer on the cartridge frit

h) Soak for one minute

i) Remove waste vial and discard methanol

From this point on do not let the cartridge dry out otherwise repeat at step f)

j) Add 30 mL of DI water to the cartridge

k) Draw the water through the cartridge to waste

4) Sample Addition

a) Add the 1 liter water sample directly to the cartridge

b) Draw the sample through the cartridge under low vacuum. This may take several minutes

depending on the solids in the sample. (Note 1) Increase vacuum pressure if necessary. Do

not exceed 500 mL/minute

c) Remove the cartridge and tap any excess water from the bottom of the cartridge

d) Replace cartridge and allow to dry under full vacuum for 10 minutes

e) Remove any water remaining in the bottom support of the cartridge with a paper towel if

necessary

5) Elution

a) Prepare an extract collection vial containing about 8 mm (0.3 inch) of anhydrous sodium

sulfate

b) Place the vial in the manifold station under the cartridge

c) Rinse the water sample bottle with 10 mL of hexane

d) Add the hexane to the cartridge

e) Soak cartridge for 2 minutes

f) Turn on vacuum and slowly draw the hexane through the cartridge and into the collection vial

g) Turn off vacuum then repeat steps 5 c) – f) 2 additional times but use 10 mL of hexane

h) Do not allow the solvent to splash into the collection vial

i) Add another 10 mL of hexane to the cartridge, rinsing the bottle holder

j) Soak cartridge for 2 minutes

k) Draw the hexane through the cartridge

6) Dry the Extract

a) Remove the collection vial from the manifold and cover with a screw cap

b) Shake the extract to form a water/hexane emulsion and immediay pour the extract through a

sodium sulfate funnel or column containing approximay 40 g of anhydrous sodium sulfate

held in place with a glass wool plug or frit. Do not use filter paper

c) Collect the extract in a clean, pre-weighed vessel

d) Rinse the collection vial with hexane and add it to the sodium sulfate. This will rinse the vial

and the sodium sulfatez

7) Gravimetric analysis

a) Carefully evaporate the hexane using a TurboVap® or similar evaporator at 40° C until the

extract just reaches dryness.

Note: Do not over dry or low recoveries will result

Do not dry on a hot plate or in an oven or over dry as low recoveries may result

b) Allow to cool to room temperature in a desiccator before weighing

c) Record this weight as the mass per unit volume of oil and grease and report as HEM as mg/L

Notes

1) If very high solids are present, add glass wool to the cartridge prior to extraction to prevent clogging and improve

flow. The glass wool must be thoroughly rinsed as part of the cartridge during the elution step.

2) Stearic acid must be in solution in the spiking solution or low recoveries will result. If small crystals are present in

the spiking solution, sonicate or shake until dissolved.

DCN-118170-

[ 打印 ] [ 返回顶部 ] [ 关闭
 
销 售:86-020-36536172,36536170,36536193
传 真:86-020-36536169
手 机:13724099587
技术服务:020—36536172
在线QQ:
 
GB/T28112-2001硅橡胶中挥发性甲基硅氧烷含量测定
GBT-18614-2012 七氟丙烷气相法测定
盐酸头孢替安有关物质分析谱图
EPA1664—己烷提取物的检测
HPLC法测定谷物中黄曲霉毒素B1、G1、B2、G2含量
液相色谱法分析多环芳烃
HP-SCX强阳离子交换色谱柱使用手册
气相色谱分析概述
气相色谱分析的特点及其应用

广州分分快三下载科学仪器有限公司 版权所有 总访问量:931778 地址:广东省广州市天河区元岗街道元岗路天河慧通产业广场A1栋六楼1604(总部) 邮编:510515
联系人:钟生 邮箱:fighter31668@163.com
GoogleSitemap 网址:www.gzalson.com 技术支持:化工仪器网 ICP备案号:
网站关键词:inertsil 色谱柱,inertsustain 色谱柱,分分快三下载液相色谱仪,kromasil,hypersil ,分分快三下载

推荐收藏该企业网站